IT Power Supply Systems

Automatic detection of electrical ground faults and prompt location

TECHNICAL DATA SHEET PHOTO GALLERY
GENERAL INFORMATION

Insulated ground system, with automatic detection of electrical ground faults and their prompt location, for critical hospital areas.

International and local standards require insulated voltage power systems for hospital critical areas, with immediate ground fault detection, allowing the system to continue online in the event of a simple fault condition.

The electrical safety insulation panel provides an insulated power supply system for critical areas, such as operating rooms, ensuring that electrical ground faults are detected and located quickly and automatically, complying with applicable standards.

MAIN CHARACTERISTICS

Automatic switching
Automatic input power supply system for emergency line or lines, with a cutting time of less than 90ms. With optional manual bypass.

Isolation transformer for medical use
Single or three phase transformer, with power from 1 to 10KVA, and 7MΩ insulation resistance

Insulation Monitor
Control of leakage current according to the limits of the hospital critical unit environment. Measurement of transformer temperature, load and control of alarm history

Circuit break
Number of circuit breakers according to outputs required for the area

Modbus connection
Transmission by serial communication between the equipment that is part of the electrical safety insulation panel

TECHNICAL DATA SHEET